Meas. ranges: -50.0°C to 760°C /-58.0°F to 1400°F; Sample rate: Less than 1 second; Basic accuracy: ±2% of reading or ±2°C/±4°F; Resolution: 0.1°C/1°C; Over range indication: LCD will show "OL" and "-OL"; Polarity: Automatic (no indication for positive polarity); Minus (-) sign for negative polarity; Emissivity: 0.95 fixed value; Field of view: D/S = 10:1; Diode laser: Output <1mW, Wavelength 630~670nm, class 2 (II) Laser product; Spectral response: 6~14μm; Power off: Automatic shut off after 7 seconds, approx.; Operating temp.: 0°C to 50°C (32°F to 122°F); Storage temp.: -20°C to 60°C (-4°F to 140°F)
AKTAKOM ATE-2523 is capable of non-contact (infrared) temperature measuring at the touch of a button. The built-in laser pointer increases target accuracy while the backlight LCD and handy push-buttons combine for convenient, ergonomic operation.
Features
Precise non-contact measurements
Built-in laser pointer
Automatic selection range and Resolution to 0.1° or 1°
ºC/ºF switchable button
Automatic Data Hold & Auto power off
The meter at 10 inches away measure 1 inch target
Backlit LCD display
MAX/MIN temperature measurement Function
Overrange indication and automatic power off
Unique flat surface, mini modern housing design
Accuracy
Accuracy is given at 18°C to 28°C (64°F to 82°F), less than 80%RH
50.0°C~-20.0°C (-58.0°F~-4.0°F): ±5°C/±9°F
-20.0°C~200.0°C (-4.0°F~200.0°F): ±2% of reading or ±2°C/±4°F
200.0°C~760°C (-200°F~1400°F): ±2.5% of reading or ±2°C/±4°F
Specifications
Meas. ranges: -50.0°C to 760°C/-58.0°F to 1400°F
Sample rate: Less than 1 second
Basic accuracy: ±2% of reading or ±2°C/±4°F
Resolution: 0.1°C/1°C
Overrange indication: LCD will show "OL" and "-OL"
Polarity: Automatic (no indication for positive polarity); Minus (-) sign for negative polarity.
Emissivity: 0.95 fixed value
Field of view: D/S = 10:1
Diode laser: Output <1mW, Wavelength 630~670nm, class 2 (II) Laser product
Spectral response: 6~14μm
Power off: Automatic shut off after 7 seconds, approx.
Hold the meter by its Handle Grip and point it toward the surface to be measured
Pull and hold the Trigger to turn the meter on and begin testing. The display will light if the battery is good. Replace the battery if the display does not light
While measuring, the SCAN display icon will flicker in the upper left hand corner of the LCD
While continuing to pull the Trigger:
Push the Laser button to turn on the laser pointer. When the laser is on the laser icon will appear on the LCD over the temperature. Aim the red beam approximately a half inch above the point of test (pressing the Laser button again turns the laser off).
Select the temperature units (°C or °F) using the °C/°F buttons.
Push the Backlight key to turn on the LCD backlighting function.
Push the mode key to enter the max temperature or min temperature measurement
Release the Trigger and the HOLD display icon will appear on the LCD indicating that the reading is being held
The meter will automatically power down after approximately 7 seconds after the trigger is released.
How can I measure temperature with ATE-2523 pyrometer?
Pull and hold the Trigger to turn the meter on and begin testing. The display will light if the battery is good. Replace the battery if the display does not light. While measuring, the SCAN display icon will flicker in the upper left hand corner of the LCD.
Release the Trigger and the HOLD display icon will appear on the LCD indicating that the reading is being held.
The meter will automatically power down after approximately 7 seconds after the trigger is released.
Most (90% of typical applications) organic materials and painted or oxidized surfaces have an emissivity of 0.95 (pre-set in the unit).
Inaccurate readings will result from measuring shiny or polished metal surfaces. To compensate, cove the surface to be measured with masking tape or flat black paint. Allow time for the tape to reach the same temperature as the material underneath it. Measure the temperature of the tape or painted surface.